At the initial fitting, the first step is to run the
Feedback Optimization test to activate the
phase canceller. This test will take
approximately 15 seconds per ear.
To run the Feedback Optimization test:
1. Ensure that ambient noise is at a
minimum. If the ambient noise level is too
high during the test, U:fit will prompt you
to reduce ambient noise levels.
2. Click either the Right Ear or Left Ear
button to start the test. Test progression
can be seen on the status bar at the
bottom of the screen. To stop the test,
click Stop.
If the test ran successfully, the hearing aid’s
gain limits are updated on the graph (black line) and the phase canceller is enabled.
High ambient noise levels may affect the Feedback Optimization results. If this is the case,
you will be notified at the end of the test. You may choose to accept the results or reduce
ambient noise levels and conduct the test again.
Curve description:
Note: The phase canceller is not enabled until the feedback optimization test has been run.
For slim tube fittings the speech enhancement is not enabled until the feedback
optimization test has been run.
Black line Feedback threshold (max gain achieved without feedback)
Green line Target gain
Red or Blue line Aided gain
Grey line Hearing instrument limit
Element Programming Guide